Total Dissolved Gas and Oxygen Monitor
Model TBO-F


Sensors
Barometric Pressure
Temperature
Total Gas Pressure-Pt
Oxygen-pO2
Derived Parameters
Delta-P
Pt-pO2 (= N2 + Ar)
% Saturation - TDGP
% Saturation 02



SPECIFICATIONS: Model TBO-F



FEATURES OF TBO-DL6 METERS:
Introducing the Common Sensing TBO-DL6 Series


SPECIFICATIONS: TBO-DL6 "-L", "-F" & "-H" Models:

CASE TYPES:
TBO-DL6-F: Pelican waterproof
TBO-DL6-H: Wall or Pedestal mounted Industrial windowed case for fixed sites
TBO-DL6-L: Bone-colored plastic bench top case

POWER:
Internal: 1.23 amp hour rechargeable gel cell
External: 9 to 24 VAC or VDC @ 200mA or greater (wall charging module included)

SENSOR PARAMETERS:
Barometer - barometric pressure in mmHG (+/- 2 mmHG)
Temperature - probe temp in degrees Celsius (+/- .2 C)
Total Dissolved Gas Pressure - probe TDGP or "Pt" in mmHG (+/- 2 mmHG)
Oxygen - probe partial pressure pO2 in mmHG (+/-3 mmHG)
Relative Humidity - atmospheric RH in percent (+/- 5% from 5% to 95%)

DERIVED PARAMETERS:
Delta-P - TDGP - BAR in mmHG (+/- 3mmHG)
Pt -pO2 - TDGP - pO2 (Remaining Gas Pressure after O2 is subtracted, nominally equal to nitrogen/Argon after saturated water vapor is subtracted)
% SATN TDGP - percentage ratio of TDGP or Pt to Barometric Pressure (Pt / BAR * 100)
% SATN O2 - percentage ratio of pO2 to expected pO2 (based on Barometric Pressure and corrected for RH when calibrated in air) (pO2 / expected pO2 * 100)
Compensation Depth - the depth below which bubble formation or growth will not occur

DISPLAY:
4-line, 20-character dot-matrix LCD

COMMUNICATION:
RS232 serial (19.2Kb)
PC Printer parallel port

DATA MEMORY:
Total Maximum - 1,000KB.

DATA OUTPUT FORMAT:
Microsoft Excel compatible (ASCII data Tab delimited)

PC SUPPORT SOFTWARE:
A PC based communications program is provided to permit -
Viewing TBO-DL6 current data output
Interaction with the TBO-DL6 Setup Menu
Offload data that has been logged
Install new software program in TBO-DL6. NOTE: Logging can be
started or stopped without the PC using the "LOG" button on the TBO-DL6

PHYSICAL:
TBO-DL6-F - L - , W - , H - , Weight - 3 lbs
TBO-DL6-H - L - , W - , H - , Weight - 3 lbs
TBO-DL6-L - L - , W - , H - , Weight - 3 lbs



CALIBRATED PROBES FOR:
(1)Temperature, (2)Total Dissolved Gas Pressure & (3)Oxygen

The Probe, which uses the same internal and external Housing as all TB-model probes, has a newinternal board, pressure sensor and temperature sensor with more rapid time respnses (20 secondsfor Temperature and 4-5 minutes for PT.) It uses OXYGUARD O2 sensor for oxygen. Upgrading ofexisting equipment can be done for 25% discount off list prices.

Any length cable can be used with no need for recalibration.

The TBOCP can be used alone with any system reading analog voltage from 0.0 to 1.5 volts. Mostsuch systems give the user a choice of 0-100 volts, 0-50 volts and 0-10 volts DC. In mostsituations 0-1.0 volts will be adequate unless dissloved gas pressures of greater than 1000 mmm Hgare expected.


PROBE SUPPLY: 12 Volts DC (unregulated)

Cable length (standard) 20 feet; extra length @ 1.80



DISSOLVED GAS EXCHANGE MEMBRANE: Dimethyl Silicon tubing (course or fine) wound on ourstandard cartridge which allows quick replacement (<1.0 minute) when necessary and comes with perforated metal shield to minimize membrane breakage from scouring of particulates and biological action.